nCPAP driver unit for neonates and premature infants
CPAP driver
Technical specifications
- Universal
driver with pneumatic gas blender for oxygen and air
- Setting
range 21% to 100% oxygen - step less
- Automatic
alarm system for drop in air or oxygen pressure
- Independent
from electrical power
- Pressure
of oxygen and air supply: 3,5 to 6 bar.
- Pressure
difference of air and oxygen supply < 1 bar
- Integrated
flow meter with needle valve, range 0 to 16 L / min.
- Flow
outlet connector, standard connection 15 and 22 mm
- Integrated
oxygen sensor for FiO2 measurement
- Type
of oxygen sensor: OOM102 (galvanic),
- Mechanical
Blender independent from electronic and monitor functions
- The
electronic system is completely battery driven and recharged by an ext. Power
supply
- Operating
time in battery-powered mode is approx. 5 hours (for transportation)
- Patient
data output by RS 232
sindi
REF 1080 sindi
- Display of nCPAP pressure in analogue curve and digital form
- Display of oxygen concentration in bargraph and digital form
- Display spontaneous breathing frequency in blinking and digital form
- Display of flexible setting upper and lower alarm for nCPAP pressure
- Reset for alarms - mute 2 minutes or quit
- Disconnection alarm for nCPAP (if pressure is lower than 1,5cm H2O)
- Integrated service software
- Batterie operating for about 6 hrs. with charging control
Available accessories for sindi driver:
- Calibration set
- Tools for maintenance
- Maintenance video
- User manual video
- Update kit for software
Gas blender for O2 and Air
REF 1085 medin Blender
- Setting range 21% to 100% Oxygen - step less
- Integrated Flow meter range 0 to 16 L/min
- Alarm system for drop gas supply pressure (Air and O2)
- Inlet pressure for Air and Oxygen 3,5 to 5 bar
- Outlet connector tapped outlet connector or on request
- Standard connection 15mm to 22mm
- Mounting for rail.
Please note: For a complete CPAP system, additional water column REF 1050 is necessary.
Easy gas blender for O2 and Air
REF 1085 EASY 15
-
Setting range 21% to 100% Oxygen - step less
- Integrated Flow meter range 0 to 15 L/min
- Alarm system for drop gas supply pressure (Air and O2)
- Inlet pressure for Air and Oxygen 3,5 to 5 bar
- Outlet connector tapped outlet connector or on request
- Standard connection 15mm to 22mm
- Mounting for rail
Please note: for a complete CPAP system, additional water column REF 1050 is necessary.
Water column
REF 1050
Water column
Neo gas blender
REF 1085 EASY 3
- Setting range 21% to 100% Oxygen – step less
- Integrated Flow meter range 0 to 3 L/min
- Alarm system for drop gas supply pressure (Air and O2)
- Inlet pressure for Air and Oxygen 3,5 to 5 bar
- Outlet connector tapped outlet connector or on request
- Standard connection 15mm to 22mm
- Mounting for rail.
Pole with mounting
REF 5001, REF 5002
REF 5001 pole with mounting REF 5002
Pole for mounting and with bottle holder
REF 5011