nCPAP driver unit for neonates and premature infants

CPAP driver

Technical specifications

  • Universal driver with pneumatic gas blender for oxygen and air
  • Setting range 21% to 100% oxygen - step less
  • Automatic alarm system for drop in air or oxygen pressure
  • Independent from electrical power
  • Pressure of oxygen and air supply: 3,5 to 6 bar.
  • Pressure difference of air and oxygen supply < 1 bar
  • Integrated flow meter with needle valve, range 0 to 16 L / min.
  • Flow outlet connector, standard connection 15 and 22 mm
  • Integrated oxygen sensor for FiO2 measurement 
  • Type of oxygen sensor: OOM102 (galvanic), 
  • Mechanical Blender independent from electronic and monitor functions
  • The electronic system is completely battery driven and recharged by an ext. Power supply
  • Operating time in battery-powered mode is approx. 5 hours (for transportation)
  • Patient data output by RS 232

sindi

sindi REF 1080 sindi
  • Display of nCPAP pressure in analogue curve and digital form
  • Display of oxygen concentration in bargraph and digital form
  • Display spontaneous breathing frequency in blinking and digital form
  • Display of flexible setting upper and lower alarm for nCPAP pressure
  • Reset for alarms - mute 2 minutes or quit
  • Disconnection alarm for nCPAP (if pressure is lower than 1,5cm H2O)
  • Integrated service software
  • Batterie operating for about 6 hrs. with charging control
Available accessories for sindi driver:
  • Calibration set
  • Tools for maintenance
  • Maintenance video
  • User manual video
  • Update kit for software
 

Gas blender for O2 and Air

Gas blender for O2 and Air REF 1085 medin Blender
  • Setting range 21% to 100% Oxygen - step less
  • Integrated Flow meter range 0 to 16 L/min
  • Alarm system for drop gas supply pressure (Air and O2)
  • Inlet pressure for Air and Oxygen 3,5 to 5 bar
  • Outlet connector tapped outlet connector or on request
  • Standard connection 15mm to 22mm
  • Mounting for rail.
Please note:
For a complete CPAP system, additional water column REF 1050 is necessary.
 

Easy gas blender for O2 and Air

Easy gas blender for O2 and Air REF 1085 EASY 15
  • Setting range 21% to 100% Oxygen - step less
  • Integrated Flow meter range 0 to 15 L/min
  • Alarm system for drop gas supply pressure (Air and O2)
  • Inlet pressure for Air and Oxygen 3,5 to 5 bar 
  • Outlet connector tapped outlet connector or on request
  • Standard connection 15mm to 22mm
  • Mounting for rail
Please note:
for a complete CPAP system, additional water column REF 1050 is necessary.
 

Water column

Water column REF 1050
Water column
 

Neo gas blender

Neo gas blender  REF 1085 EASY 3
  • Setting range 21% to 100% Oxygen – step less
  • Integrated Flow meter range 0 to 3 L/min
  • Alarm system for drop gas supply pressure (Air and O2)
  • Inlet pressure for Air and Oxygen 3,5 to 5 bar
  • Outlet connector tapped outlet connector or on request
  • Standard connection 15mm to 22mm
  • Mounting for rail.
 